• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT POINT OF CARE I-STAT CREA CARTRIDGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT POINT OF CARE I-STAT CREA CARTRIDGE Back to Search Results
Catalog Number 03P84-25
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Codes Peripheral Vascular Disease (2002); Claudication (2550)
Date of Event 05/10/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Apoc incident # (b)(4) apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Event or Problem Description
On (b)(6) 2019, abbott point of care was contacted by a customer regarding i-stat crea cartridges that yielded a suspected discrepant creatinine results on a (b)(6) year old female patient with peripheral vascular disease with claudication.Both i-stat results were consistent at 2.4 mg/dl from a venous line.A repeat on the alternate method was 0.5 mg/dl via venipuncture.There was no additional patient information available at the time of this report.Return product is available for investigation.(b)(6).There are no injuries associated with this event.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
 
Additional Manufacturer Narrative
Apoc incident #: (b)(4).The investigation was completed on 06/10/2019.A review of the device history record confirmed the lot passed finished goods release criteria.Retained and returned cartridge testing met the acceptance criteria found in q04.01.003 rev.Ad, appendix 1- product complaint level 2 and level 3 investigation procedure.No deficiency has been identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT CREA CARTRIDGE
Common Device Name
CREA CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
MDR Report Key8615333
Report Number2245578-2019-00128
Device Sequence Number2074238
Product Code CGL
UDI-Device Identifier10054749000125
UDI-Public10054749000125
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K973292
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/20/2019
Device Catalogue Number03P84-25
Device Lot NumberA19021
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/21/2019
Initial Date Received by Manufacturer 05/13/2019
Supplement Date Received by Manufacturer06/10/2019
Initial Report FDA Received Date05/16/2019
Supplement Report FDA Received Date06/13/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age76 YR
-
-