• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AEYSNT INCORPORATED OMNICELL I.V.STATION; PHARMACY COMPOUNDING DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AEYSNT INCORPORATED OMNICELL I.V.STATION; PHARMACY COMPOUNDING DEVICE Back to Search Results
Model Number I.V. STATION
Medical Device Problem Codes Incorrect, Inadequate or Imprecise Result or Readings (1535); Output Problem (3005)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/06/2019
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2019 an iv bag that was prepared by an i.V.Station device had a label affixed that contained incorrect content.The label for a vancomycin preparation included a bar code for oxytocin.The mislabeled preparation was identified and there are no known adverse patient effects.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNICELL I.V.STATION
Common Device Name
PHARMACY COMPOUNDING DEVICE
Manufacturer (Section D)
AEYSNT INCORPORATED
51 pennwood place
warrendale PA 15086
MDR Report Key8615497
Report Number3011278888-2019-00057
Device Sequence Number14038228
Product Code NEP
UDI-Device Identifier08056477790007
UDI-Public(01)08056477790007(11)180108
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberI.V. STATION
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/06/2019
Initial Report FDA Received Date05/16/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-