| Model Number |
9004 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Reocclusion (1985)
|
| Date of Event |
02/01/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Analysis: the samples were not returned from the user facility; therefore, device evaluations are unable to be performed.A lot history review revealed two additional reocclusion complaints were associated with this lot at the complaint was created.A review of the device history record (dhr) indicates the lot was manufactured to specification.Conclusion: the actual samples were not returned for evaluation.The dhr found nothing to indicate a manufacturing related cause for this event.The investigator assessed the event was not related to the study devices or procedure.Based on the instructions for use (ifu), the occurrence of reocclusion is an inherent risk of any pta procedure, and has been reported in clinical trials of drug coated balloons.If additional information becomes known to the manufacturer, a supplemental report will be provided with all relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported through a clinical registry that during the index procedure, two lutonix percutaneous transluminal drug coated balloon (dcb) dilatation catheters were used to treat the target lesions located in the left mid superficial femoral artery (sfa) and left mid popliteal artery.Approximately 13 months after the index procedure, the patient¿s target lesions were reportedly reoccluded via duplex ultrasound (dus) imaging.The registry's independent evaluator assessed that both target lesions were patent at this time.Approximately 29 months after the index procedure, a revascularization was performed on target lesion 1, left mid popliteal artery.The investigator assessed that the event was not related to the study devices or procedure.The samples were discarded by the user facility and are not available for further evaluation.No adverse patient effects were reported.This is one of two products involved with the reported event and the associated manufacturer¿s report number is 3006513822-2018-00266.
|
| |
|
Search Alerts/Recalls
|