|
Device passed the functional test, including the displacement test, rewind, basic occlusion test, occlusion test, prime test, excessive no delivery test and delivery accuracy test at 0.08740 inches.The stop (idle) current and run current measurement tests are within specification.Device also passed self test, off no power alarm test and unexpected restart error test.Device uploaded properly using carelink.Device communicated properly with the glucose sensor simulator and the minilink transmitter.Calibrated the sensor transmitter.During testing the test value of 135 mg/dl was properly displayed on the pump sensor graph screen.No pump/sensor transmitter communication anomaly, calibration anomaly or lost sensor alarm noted during testing.Device had cracked battery tube threads, cracked reservoir tube lip and a stained end cap sticker.(b)(4).
|