| Catalog Number |
4100355000 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
04/25/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The device had a bit broken off inside during a service evaluation at the manufacturer facility. there were no adverse consequences related to this event.
|
| |
|
Event or Problem Description
|
|
The device had a bit broken off inside during a service evaluation at the manufacturer facility. there were no adverse consequences related to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: follow-up report submitted to document device evaluation results.
|
| |
|
Search Alerts/Recalls
|