• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION POLARIS ULTRA; STENT, URETERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION POLARIS ULTRA; STENT, URETERAL Back to Search Results
Model Number M006193131090
Medical Device Problem Code Unsealed Device Packaging (1444)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/25/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.It was reported the device was not used past its expiry.(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a polaris ultra stent was to be used in a procedure performed on (b)(6) 2019.According to the complainant, during unpacking, it was noticed that the packaging of a polaris ultra stent was open and the sterility of the device was compromised.The procedure was successfully completed with another polaris ultra stent.There was no serious injury, nor were there any adverse patient effects reported as a result of this event.
 
Additional Manufacturer Narrative
The device was returned with its original opened pouch.A visual analysis of the retuned opened pouch found that the sealed marks were evidence that the selaed procesds was properly performed.No visual defects were observed on the stent and the pouch.Basd on product analysis, and all compiled information, the most probable cause of this event is no problem detected since the complaint included a returned device review (visual, physical, and performance testing) which showed no evidence of either the alleged issue or any defect which could have contributed to the event.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications at the time of release for distribution.
 
Event or Problem Description
It was reported to boston scientific corporation that a polaris ultra stent was to be used in a procedure performed on (b)(6) 2019.According to the complainant, during unpacking, it was noticed that the packaging of a polaris ultra stent was open and the sterility of the device was compromised.The procedure was successfully completed with another polaris ultra stent.There was no serious injury, nor were there any adverse patient effects reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARIS ULTRA
Common Device Name
STENT, URETERAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key8619800
Report Number3005099803-2019-02565
Device Sequence Number2039320
Product Code FAD
UDI-Device Identifier08714729762904
UDI-Public08714729762904
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K010002
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/07/2021
Device Model NumberM006193131090
Device Catalogue Number193-131-09
Device Lot Number0022939917
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/24/2019
Initial Date Received by Manufacturer 04/25/2019
Supplement Date Received by Manufacturer06/26/2019
Initial Report FDA Received Date05/17/2019
Supplement Report FDA Received Date07/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-