| Medical Device Problem Code |
Structural Problem (2506)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2019, the reporter contacted animas, alleging a casing/condition (cracked/damaged casing) issue; a cracked battery compartment.There is no indication the alleged product issue caused or contributed to an adverse event.This complaint is being reported because the alleged malfunction can result in a delay in treatment or long term cessation in delivery if the damage impacts the power circuit or cartridge compartment.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device has been returned and evaluated by product analysis on 04-jun-2019 with the following findings:.During investigation, the battery compartment was found to be cracked.
|
| |
|
Search Alerts/Recalls
|