• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDPLAST MEDICAL, INC. REZUM; UNIT, ELECTROSURGICAL ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDPLAST MEDICAL, INC. REZUM; UNIT, ELECTROSURGICAL ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Model Number D2201
Device Problems Material Fragmentation (1261); Appropriate Term/Code Not Available (3191)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/19/2019
Event Type  malfunction  
Event Description
It was reported that during convective radiofrequency water vapor thermal therapy treatment of the prostate, while delivering the 10 th treatment, a plastic flaking off was able to be seen inside the patient.The material was flushed out of the patient's body.The procedure was completed utilizing the same device without clinical consequences to the patient.
 
Manufacturer Narrative
The returned device was disassembled and the fragmented piece was identified to come from the silicon duck bill that the scope enters through in the back of the device.This was most likely caused by the scope being pushed in with too much force or pushed in at a non-linear angle.A review of the device history record confirmed that the device met all material, assembly and performance specifications prior to release.A review of the device ifu and operators manual was completed and did not reveal any evidence of device misuse, off-label use, or failure to follow instructions.Based on the analysis results, the information provided is not sufficient to assign a cause, therefore, the investigation conclusion code is cause not established.
 
Event Description
It was reported that during convective radiofrequency water vapor thermal therapy treatment of the prostate, while delivering the 10th treatment, a plastic flaking off was able to be seen inside the patient.The material was flushed out of the patient's body.The procedure was completed utilizing the same device without clinical consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REZUM
Type of Device
UNIT, ELECTROSURGICAL ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
MEDPLAST MEDICAL, INC.
5079 33rd street se
grand rapids MI 49512
MDR Report Key8620703
MDR Text Key145406447
Report Number2937094-2019-60634
Device Sequence Number1
Product Code KNS
Combination Product (y/n)N
PMA/PMN Number
K180237
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 10/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/17/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2019
Device Model NumberD2201
Device Catalogue NumberD2201
Device Lot Number2018110702
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/09/2019
Date Manufacturer Received10/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-