Concomitant medical products: erbe electrosurgical generator cook acrobat 2 calibrated tip wire guide, awg2-35-260.Investigation evaluation: our evaluation of the product said to be involved determined the cutting wire securing component located near the distal end of the sphincterotome has disconnected from the catheter.The cutting wire was intact and remained securely attached to the sphincterotome at the proximal end.However, due to the catheter and securing component disconnection, the distal end of the cutting wire was no longer connected to the sphincterotome catheter at the distal end.The securing component was intact, therefore no part of the device is missing.The cutting wire does not show evidence of cautery application.A visual examination of the catheter showed fraying approximately 112 cm and 116 cm from the distal end.A product discrepancy or anomaly that could have contributed to this reported occurrence was not observed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.Separation of the cutting wire securing component and the catheter can occur if the tip of the sphincterotome is over flexed.The instructions for use caution the user: "do not over flex or bow tip beyond 90 degrees, as this may damage or cause cutting wire to break." this type of damage can occur if the distal end of the catheter is shaped manually.This sphincterotome catheter is precurved and is provided with a precurved stylet in the distal tip of the catheter.This obviates the need for manual formation.The instructions for use contain the following comment: ¿note: do not apply manual pressure to tip or cutting wire of sphincterotome in an attempt to influence orientation, as this may result in damage to device.¿ other factors that can contribute to separation of the cutting wire securing component and the catheter include manipulating the handle with the catheter in a coiled position or with the precurved stylet inside the cannulating tip.The instructions for use advise the user: "upon removing device from package, uncoil and straighten sphincterotome.Carefully remove precurved stylet from cannulating tip." the instructions for use contain the following comment: "note: do not exercise handle while device is coiled or precurved stylet is in place, as this may cause damage to sphincterotome and render it inoperable." prior to distribution, all fusion omni-tomes are subjected to a visual inspection and functional test to ensure device integrity.The functional test includes bowing the sphincterotome to ensure the distal end responds to handle manipulation.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
|
During an endoscopic retrograde cholangiopancreatography (ercp), the physician used a cook fusion omni-tome sphincterotome.The wire [catheter] from the sphincterotome was frayed [difficult stripping].The device was removed from the patient and a new sphincterotome was opened.There was] no injury to the patient.There was no reportable information at this time.The device was received on 04/23/2019 and the evaluation determined that the cutting wire securing component separated from the catheter.Our attempts to collect specific patient outcome information were unsuccessful.Because the complainant stated that there was no injury to the patient, we can conclude that this information does not reasonably suggest the patient was adversely impacted.
|