|
Evaluation summary: the product was not returned.A picture was provided by the customer for analysis.The picture showed damage, probably similar to a burn, to a patient's skin.The device was not visible in the photo.The picture showed damage to the skin of the patient.The device was not visible in the provided photo and was not received, and so no product evaluation could be conducted.No defects with the device were identified.The provided information was inadequate to determine a root cause for the reported event.The instructions for use(ifu) states, select a well vascularized, convex area in close proximity to the surgical site for patient return electrode application.Avoid scar tissue, bony prominences, excessive adipose tissue, and areas where fluid may pool.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|