• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PIN TRL LNR 10D 620DX36ID; HIP INSTRUMENTS : ACETABULAR TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PIN TRL LNR 10D 620DX36ID; HIP INSTRUMENTS : ACETABULAR TRIALS Back to Search Results
Catalog Number 221836162
Medical Device Problem Codes Break (1069); Scratched Material (3020)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/29/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that they attempted to use 36x62x+4 10* liner trial.Set screw and washer fell out.No broken pieces left in patient, no surgery delay.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Examination of the returned instrument confirmed the reported observation.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : null.Device history batch : null.Device history review : null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIN TRL LNR 10D 620DX36ID
Common Device Name
HIP INSTRUMENTS : ACETABULAR TRIALS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8621477
Report Number1818910-2019-93940
Device Sequence Number2182615
Product Code LXH
UDI-Device Identifier10603295101314
UDI-Public10603295101314
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number221836162
Device Lot NumberSZ1208
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/08/2019
Initial Date Received by Manufacturer 04/29/2019
Supplement Date Received by Manufacturer06/24/2019
Initial Report FDA Received Date05/17/2019
Supplement Report FDA Received Date06/24/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age49 YR
-
-