ST. JUDE MEDICAL, INC. INQUIRY AFOCUSII¿ (4 F DOUBLE LOOP) CATHETER 20 ELECTRODE, 7F; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number IBI-87008
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 05/14/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event or Problem Description
During an atrial tachycardia procedure, the catheter became lodged in the patient and could not be removed.Following mapping in the right atrium, fluoroscopy revealed the catheter would not move and could not be removed from the inferior cava.A tee did not determine where the catheter was located, and the patient was transferred to surgery to extract the catheter via the subclavian vein.The patient is recovering in the uci.
 
Event or Problem Description
It is possible that the device was trapped in fibers between the inferior vena cava and the lateral wall.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INQUIRY AFOCUSII¿ (4 F DOUBLE LOOP) CATHETER 20 ELECTRODE, 7F
Common Device Name
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
2375 morse ave
irvine CA 92614
MDR Report Key8623564
Report Number2030404-2019-00035
Device Sequence Number2257316
Product Code DRF
UDI-Device Identifier05414734309929
UDI-Public05414734309929
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K101623
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2021
Device Model NumberIBI-87008
Device Lot Number6602300
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/14/2019
Supplement Date Received by Manufacturer05/22/2019
Initial Report FDA Received Date05/20/2019
Supplement Report FDA Received Date05/24/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age31 YR
Patient Weight68
-
-