The patient reported that their generator is ¿going on and off.¿ the patient was seen by their neurologist regarding this report of erratic stimulation, and the neurologist believed that the device was ¿malfunctioning.¿ the neurologist mentioned that the patient all of sudden was having an increase in seizures after being seizure free.The physician additionally reported that it was observer that generator was set at 1% duty cycle; however, the physician denies purposefully setting the duty cycle to 1%.A system diagnostics test was performed at the patient¿s appointment and confirmed that the device was functioning properly.The doctor turned off the generator, as the patient stated it was bothering her and referred the patient to the neurosurgeon for battery replacement.An implant card was received confirming that the patient's generator was explanted and replaced due to battery depletion and "irregular stimulation".An image of the parameters and system diagnostics test from the patient¿s appointment was later received.Reviewing the patient¿s parameters it was noticed that the patient was programmed to a 29% duty cycle; however, the programmer falsely displayed a 1% duty cycle.There is a known display error that would explain why a 1% would show after system diagnostics.Any time system diagnostics is performed in-session and the user navigates back to the parameters screen, all values will display correctly (including on time and off time), except for the calculated duty cycle, which will show 1%.This display occurs because the system diagnostics will program the generator to this duty cycle to run diagnostics prior to programming them back to the intended cycle.In these cases the generator is operating at intended duty cycles settings and re-interrogation will resolve the display error.No other relevant information has been received to date.
|
|
Further follow up with the physician's office confirmed that the patient's increased seizures were above pre-vns levels and that the believed root cause for the increased seizures was a depleted battery.It was noted that to correct the issue of inconsistent stimulation the patient had to undergo battery replacement.The facility provided programming data, which did not support a depleted battery condition as of 05/13/2019.Additionally, the patient's vns generator had been programmed off since 04/25/2019 due to the reported erratic stimulation.No other relevant information has been received to date.
|