• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. META ANTERIOR DROP; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. META ANTERIOR DROP; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 71654501
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/27/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during surgery, device broke during impaction.Nothing got into the patient.No harm to patient or staff.2 minutes delay in procedure due to incident.Smith & nephew back up was available for use if needed.
 
Additional Manufacturer Narrative
A visual inspection of the returned device confirmed the stated failure mode.The device fractured into two pieces and both pieces were returned.The device was manufactured in 2017 and exhibits signs of extensive wear/usage.The fracture surface shows multiple regions of crack initiation.This was likely caused by bending, torsional, or impact mechanical overload.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.We will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
META ANTERIOR DROP
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8626425
Report Number1020279-2019-01987
Device Sequence Number1858443
Product Code HSB
UDI-Device Identifier03596010585820
UDI-Public03596010585820
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K122170
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71654501
Device Lot Number17KSP0015
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/13/2019
Initial Date Received by Manufacturer 04/27/2019
Supplement Date Received by Manufacturer06/11/2019
Initial Report FDA Received Date05/21/2019
Supplement Report FDA Received Date06/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-