|
The customer reported via phone call that they went hospital due to high blood glucose and diabetic ketoacidosis on unknown date.Customer¿s blood glucose was 446 mg/dl in the hospital.Customer had blood glucose of 40 to 100.Customer had symptoms such as nausea, vomiting abdominal pain, difficulty breathing.Customer was treated that with insulin drip.Customer stated that the buttons were hard to push.Insulin pump was scratched.The insulin pump will not be returned for analysis.
|
|
Device passed the functional test, including the displacement test, rewind, basic occlusion test, occlusion test, prime/a33 test, excessive no delivery test and delivery accuracy test at 0.08750 inches.No under delivery anomaly or over delivery anomaly noted.The stop current and run current measurement tests are within specification.Device also passed self test, off no power alarm test and a21 error test.The no delivery alarm functions properly during the basic occlusion test and occlusion test.No unexpected no delivery alarm noted during testing.Device uploaded properly using carelink.Device had a stripped battery cap at coin slot.Device had cracked case at display window corner, cracked battery tube threads, broken belt clip slot, cracked reservoir tube, cracked reservoir tube window, scratched reservoir tube window and cracked reservoir tube lip.The test p-cap and reservoir does lock in place in the reservoir compartment.(b)(4).
|