• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANIMAS LLC ANIMAS VIBE; INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS LLC ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Medical Device Problem Code Retraction Problem (1536)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event or Problem Description
On (b)(6) 2019, the reporter contacted animas alleging the pump would not perform the rewind step and the piston rod inside the cartridge compartment was not retracting.There is no indication the alleged product issue caused or contributed to an adverse event.No further information was available.Animas customer support made several attempts to follow-up with the reporter to perform troubleshooting of the issue; however, the reporter did not respond to the follow-up attempts.This complaint is being reported because the insulin delivery could be affected.
 
Additional Manufacturer Narrative
Device evaluation: the device has been returned and evaluated by product analysis on 08-oct-2019 with the following findings:.During investigation, the complaint regarding rewind not retracting was unable to be duplicated.There was no evidence of rewind alarms or warnings were observed in the black box or alarm history.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS LLC
965 chesterbrook blvd
wayne PA 19087
MDR Report Key8627114
Report Number2531779-2019-03227
Device Sequence Number2308929
Product Code OYC
UDI-Device Identifier10840406100051
UDI-Public10840406100051
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P130007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2019
Was the Report Sent to FDA? Yes
Device Age47 MO
Initial Date Received by Manufacturer 05/16/2019
Supplement Date Received by Manufacturer05/16/2019
Initial Report FDA Received Date05/21/2019
Supplement Report FDA Received Date11/05/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age77 YR
-
-