• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH RESTORATION ADM X3 INS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH RESTORATION ADM X3 INS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS Back to Search Results
Catalog Number 1236-2-244
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Inflammation (1932); Injury (2348); Reaction (2414)
Date of Event 04/24/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
 
Event or Problem Description
As reported: "cup, head, liners removed due to metallosis caused by malpositioned implants resulting in impingement".
 
Additional Manufacturer Narrative
An event regarding metallosis caused by malpositioned implants resulting in impingement was reported.Conclusion based on the information provided there is no indication or allegation that the device reported in this investigation contributed to the event and is therefore considered concomitant.The adm liner component is a polyethylene component located within the mdm liner, the reported event is regarding metallosis caused by malpositioned implants resulting in impingement, the polyethylene adm liner component would not have been in a position to generate metal debris to cause the reported metallosis.The device also does not interface with the patients bone, it is fixed within the mdm liner and hence, could not be malpositioned.A full investigation is completed on bone interfacing components - trident tritanium shell and unknown stem which may have been malpositioned resulting in impingement and also the metal mdm liner (pr 2071533) which may have impinged with the stem resulting in metallic fretting debris within the same pi.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
As reported: "cup, head, liners removed due to metallosis caused by malpositioned implants resulting in impingement".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESTORATION ADM X3 INS
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8627319
Report Number0002249697-2019-02002
Device Sequence Number11327027
Product Code LZO
UDI-Device Identifier04546540665898
UDI-Public04546540665898
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K103233
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2020
Device Catalogue Number1236-2-244
Device Lot Number282099
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/24/2019
Supplement Date Received by Manufacturer06/26/2019
Initial Report FDA Received Date05/21/2019
Supplement Report FDA Received Date07/24/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age55 YR
-
-