• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US ALTRX +4 NEUT 40IDX60OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US ALTRX +4 NEUT 40IDX60OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS Back to Search Results
Catalog Number 122140460
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hypersensitivity/Allergic reaction (1907); Inflammation (1932); Pain (1994); Scar Tissue (2060); Tissue Damage (2104); Injury (2348); No Code Available (3191)
Date of Event 05/31/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Pinnacle litigation records received alleging elevation of blood cobalt and chromium levels, sharp pains, psoas tendinitis (negative for infection), metal reaction, trunnionosis, cicatrix, physical injury and disability, lack of mobility, weakness of the left leg, difficulty walking laterally, limited mobility and emotional distress.It was noted that during the procedure, the surgeon noted a light greenish cloudy fluid and destruction of the abductor mechanism.There was a pseudotumor due to the metal-on-metal reaction and a complete synovectomy was needed to debride the hip of metal particles.There was gross corrosion with black material at the head-neck junction as well as on the head-neck taper.The liner is poly.Doi: (b)(6) 2006 (stem), doi: (b)(6) 2011 (liner and head), dor: (b)(6) 2017, left hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTRX +4 NEUT 40IDX60OD
Common Device Name
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8627945
Report Number1818910-2019-94157
Device Sequence Number19391232
Product Code LPH
UDI-Device Identifier10603295016618
UDI-Public10603295016618
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K062148
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2017
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number122140460
Device Lot Number126522
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/02/2019
Supplement Date Received by Manufacturer06/06/2019
Initial Report FDA Received Date05/21/2019
Supplement Report FDA Received Date06/07/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
-
-