| Model Number |
U875610 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Material Rupture (1546)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
04/24/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The lot number for the device was not provided.The device history records could not be performed.The device has been returned for evaluation.The investigation is currently underway.
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure of a highly calcified lesion, the pta balloon allegedly ruptured just after inflation.The device was retrieved successfully by surgical intervention.There was no further patient injury.No further treatment was necessary.
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure of a highly calcified lesion in the femoral artery, the pta balloon allegedly ruptured and could not be removed.The detached material was retrieved successfully by surgical intervention.No further treatment was necessary.There were no further reported complications.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing review: a complete manufacturing review could not be conducted, as the lot number was not reported for this device.Investigation summary: the device was returned for evaluation.A visual inspection found bunching along the outer catheter shaft and also found that the guidewire and inner lumen had been cut at the distal end of the device and that the balloon was missing.Therefore, the investigation is inconclusive for the reported balloon rupture and inconclusive for the reported entrapment based on the returned condition of the device.Per the reported event details, it is likely that the highly calcified lesion could have contributed to the reported balloon rupture and consequently the reported entrapment.However, the definitive root cause for the reported balloon rupture or the reported entrapment could not be determined based upon the available information.It is unknown whether patient and/or procedural issues contributed to the events.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.
|
| |
|