| Catalog Number |
PND6F0701058 |
| Medical Device Problem Codes |
Collapse (1099); Device Damaged Prior to Use (2284)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
04/17/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
|
| |
|
Event or Problem Description
|
|
During preparation for a medical procedure, the hospital technician found that a neuron 6f 070 delivery catheter (neuron 070) was flattened while attempting to advance a guide wire through it.The damage to the neuron 070 was found prior to use and, therefore, it was not used in the procedure.The procedure was completed using a new neuron 070.
|
| |
|
Additional Manufacturer Narrative
|
|
Results: the neuron 6f 070 delivery catheter (neuron 070) was ovalized from approximately 103.0 through 106.0 and 108.5 cm from the hub.Conclusions: evaluation of the returned neuron 070 confirmed that the distal shaft was ovalized.This type of damage typically occurs if the peel-able sheath is not properly utilized prior to use or due to forceful handling.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
| |
|