| Catalog Number |
2296100212 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/26/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Awaiting device evaluation.A follow up report will be filed once the quality investigation is complete.
|
| |
|
Event or Problem Description
|
|
It was reported that the bur broke during a molar removal.It was further reported that the broken piece was retrieved from the patient.It was also reported that there were no delays and no adverse consequences as a result of this event.It was further reported that the procedure was completed successfully.
|
| |
|
Event or Problem Description
|
|
It was reported that the bur broke during a molar removal.It was further reported that the broken piece was retrieved from the patient.It was also reported that there were no delays and no adverse consequences as a result of this event.It was further reported that the procedure was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results indicate that the bur broke as a result of being incorrectly loaded into the attachment.The instructions for use (ifu) of handpieces associated with this device provides detailed instructions for loading the cutting accessories and warn the user against this type of use: "warning: make sure the safety line on the cutting accessory is not visible after the shield is installed.Failure to comply may allow the cutting accessory to eject from the handpiece or bend.The quality investigation is complete.
|
| |
|