Catalog Number TRF-385
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 06/15/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.The reason for reoperation is capsular contracture, baker grade iv and pain.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.This is a known potential adverse event addressed in the product labeling.
 
Event or Problem Description
Patient reported left side pain and capsular contracture, baker grade unknown.Health professional clarified as baker grade iv.Device remains implanted.
 
Additional Manufacturer Narrative
Device evaluation: analysis of the returned device identified: weight to the spec, deformation.Cloudy color in the gel observed after autoclave cycle.Base on the device analysis the final assessment is: no issues found related with the manufacturing process.
 
Event or Problem Description
Device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key8629894
Report Number9617229-2019-05034
Device Sequence Number13863951
Product Code FTR
UDI-Device Identifier10888628008328
UDI-Public10888628008328
Combination Product (Y/N)N
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2019
Reporter Type Manufacturer
Report Source consumer,health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/03/2021
Device Catalogue NumberTRF-385
Device Lot Number2966960
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/22/2019
Initial Date Received by Manufacturer 04/29/2019
Supplement Date Received by Manufacturer08/09/2019
Initial Report FDA Received Date05/21/2019
Supplement Report FDA Received Date09/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age44 YR
-
-