• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT POINT OF CARE I-STAT 6+ CARTRIDGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT POINT OF CARE I-STAT 6+ CARTRIDGE Back to Search Results
Catalog Number 03P80-25
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Information (3190)
Date of Event 05/09/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Apoc incident #: (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Event or Problem Description
On (b)(6) 2019, abbott point of care was contacted by a customer regarding i-stat 6+ cartridges that yielded a suspected a discrepant potassium result on a patient.There was no patient information available at the time of this report.Method: i-stat, result: >9, lab instrument : normal (actual result unknown).Customer has not provided collect/test times, or patient information.There are no injuries associated with this event.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
 
Additional Manufacturer Narrative
Apoc incident #: (b)(4).The investigation was completed on 07/23/2019.A review of the device history record (dhr) confirmed the cartridge lot met finished goods (fg) release criteria.Retained testing met the acceptance criteria outlined in appendix 1 of q04.01.003 rev.Ad (product complaint level 2 and level 3 investigation procedure).No deficiency has been determined for 6+ lot k18301.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT 6+ CARTRIDGE
Common Device Name
6+ CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
MDR Report Key8631147
Report Number2245578-2019-00130
Device Sequence Number11246788
Product Code JGS
UDI-Device Identifier10054749000088
UDI-Public10054749000088
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K912387
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/25/2019
Device Catalogue Number03P80-25
Device Lot NumberK18301
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/09/2019
Supplement Date Received by Manufacturer07/23/2019
Initial Report FDA Received Date05/22/2019
Supplement Report FDA Received Date07/26/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-