| Catalog Number |
03P80-25 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
05/09/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Apoc incident #: (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2019, abbott point of care was contacted by a customer regarding i-stat 6+ cartridges that yielded a suspected a discrepant potassium result on a patient.There was no patient information available at the time of this report.Method: i-stat, result: >9, lab instrument : normal (actual result unknown).Customer has not provided collect/test times, or patient information.There are no injuries associated with this event.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident #: (b)(4).The investigation was completed on 07/23/2019.A review of the device history record (dhr) confirmed the cartridge lot met finished goods (fg) release criteria.Retained testing met the acceptance criteria outlined in appendix 1 of q04.01.003 rev.Ad (product complaint level 2 and level 3 investigation procedure).No deficiency has been determined for 6+ lot k18301.
|
| |
|