(b)(4).A manufacturing record evaluation was performed for the finished device mk6373 number, and no non-conformances were identified.Additional information was requested and the following was obtained: was the information on the label on the outer box and that on the individual foils different with respect to suture size, 0 and 2-0? the information on the foil and the outer box are the exact same.What was the product code and lot number on the outer box and that the foil? same on both.What was the actual size of the suture in the foil labeled as 2-0 appear to be, 0 or 2-0? 2-0.Was the issue found upon opening a new box of product? no.Is it possible that 2 products got mixed up on the shelf? no foil and box had same info.The indication on the product reel itself inside the foil indicated it was a 2-0 not 0.Was it verified that all the packages/foils in the box were same and labeled as 2.0? no.An empty opened box and an opened sample of product code j207, lot mk6373 was returned for analysis.During the visual inspection of the opened sample, the spool had two holes punched and for product code j207 should be only one hole punched.Due to mismatch at the spool a characterization of the suture was performed and meet the requirements for a vicryl suture diameter 0.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.According to the sample condition, the assignable cause of the mix reel is due a manufacturing defect.
|