• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5530-G-509
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Injury (2348); Insufficient Information (4580)
Date of Event 04/18/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The following devices were also listed in this report: triathlon asymmetric x3 patella; cat# 5551-g-320; lot# 0y94; triathlon prim cem fxd bplt #5; cat# 5520b500; lot# stehf; triathlon cr fem comp #4 l-cem; cat# 5510f401; lot# stdlb.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.The information in this report was provided by stryker orthopaedics legal affairs department.No additional information is available at this time due to the ongoing litigation.Should additional information become available, the evaluation summary will be submitted in a supplemental report.Device not returned.
 
Event or Problem Description
It was reported through the communication of an attorney that allegedly the patient was revised due to "status post left tkr with prosthetic joint implant failure".
 
Additional Manufacturer Narrative
An event regarding revision surgery involving a triathlon insert was reported.The event was not confirmed.Method & results: -device evaluation and results: not performed as product was not returned.-clinician review: a review of medical records indicated "root cause analysis: the cause of this event cannot be determined with certainty with the information provided.The reason stated for revision of the primary implant was due to "status post left tkr with prosthetic joint implant failure.".This is a nebulous and imprecise diagnosis.The term "implant failure" is often used to describe simple failure of a total knee arthroplasty due to myriad causes including infection, instability, loosening, impingement, etc.In most cases this does not relate to actual mechanical failure of the implant such as breakage.Therefore it is impossible to assign any causality to the implant with this imprecise diagnosis." -device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the reported lot.Conclusion: a review of medical records indicated "root cause analysis: the cause of this event cannot be determined with certainty with the information provided.The reason stated for revision of the primary implant was due to "status post left tkr with prosthetic joint implant failure.".This is a nebulous and imprecise diagnosis.The term "implant failure" is often used to describe simple failure of a total knee arthroplasty due to myriad causes including infection, instability, loosening, impingement, etc.In most cases this does not relate to actual mechanical failure of the implant such as breakage.Therefore it is impossible to assign any causality to the implant with this imprecise diagnosis." the exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Event or Problem Description
No new information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON CR X3 TIBIAL INSERT
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
joann ripoli
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key8632953
Report Number0002249697-2019-02060
Device Sequence Number16732739
Product Code MBH
UDI-Device Identifier07613327050530
UDI-Public07613327050530
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2017
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2013
Device Catalogue Number5530-G-509
Device Lot NumberMERL0D
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2019
Supplement Date Received by Manufacturer04/17/2025
Initial Report FDA Received Date05/22/2019
Supplement Report FDA Received Date05/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/03/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age63 YR
Patient SexMale
-
-