| Model Number |
457445 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/01/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: product id: dtma1d4 icd; 383069 lead, implanted: (b)(6) 2019.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that one day post replacement procedure, the right atrial (ra) lead had no capture.It was noted that a chest x-ray revealed the ra lead dislodged.During lead revision, after the lead was revised the physician went to reattach the lead into the header and the wrench would not engage into the set screw.The physician could not screw the lead back into the set screw.The device was explanted and replaced, and the ra lead was explanted and replaced.No patient complications have been reported as a result of this event.
|
| |
|