Model Number G22490
Device Problem Failure to Align (2522)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.Prior to distribution, all uts precurved ultra taper sphincterotomes are subjected to a visual inspection and functional test to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
In preparation for a procedure, the user selected a cook uts precurved ultra taper sphincterotome.Upon bending the cutting wire, the sphincterotome orientation was at 9 o'clock instead of 12 o'clock.The orientation was completely off.Another of the same device was opened and used to complete the procedure.This occurred prior to patient contact; there was no impact to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UTS PRECURVED ULTRA TAPER SPHINCTEROTOME
Type of Device
KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
scottie fariole
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key8633948
MDR Text Key146481838
Report Number1037905-2019-00268
Device Sequence Number1
Product Code KNS
UDI-Device Identifier00827002224902
UDI-Public(01)00827002224902(17)220215(10)W4180445
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K901443
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/22/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/15/2022
Device Model NumberG22490
Device Catalogue NumberUTS-25M
Device Lot NumberW4180445
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/01/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/25/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-