|
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #: (b)(4).Investigation summary: the instrument associated with this report was not returned.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Neck segment is loose on broach patient status/ outcome / consequences --> no, was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: --> unknown, is the patient part of a clinical study --> unknown, ip-00576625 device property of -->none, device in possession of -->none, ip-00576626 device property of -->none, device in possession of -->none, ip-00576627 device property of -->none, device in possession of -->none, ip-00576628 device property of -->none, device in possession of -->none, ip-00576629 device property of -->none, device in possession of -->none, ip-00576630 device property of -->none, device in possession of -->none, ip-00576631 device property of -->none, device in possession of -->none, ip-00576632 device property of -->none, device in possession of -->none, ip-00576633 device property of -->none, device in possession of -->none, ip-00576634 device property of -->none, device in possession of -->none, ip-00576635 device property of -->none, device in possession of -->none, ip-00576636 device property of -->none, device in possession of -->none by checking this box i certify that all information that are known/available has been disclosed.If any new information will be made available, the additional information will be submitted through cst.--> true.
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # : (b)(4).Investigation summary = > examination of the returned instrument found this instrument to be of the old design.The root cause is attributed to design.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Complaints will be monitored under post market surveillance sep 419.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|