PENUMBRA, INC. RUBY COIL; HCG, KRD Back to Search Results
Catalog Number RBY2C1240
Medical Device Problem Codes Physical Resistance/Sticking (4012); Premature Separation (4045)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/23/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2019-01002.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the internal iliac artery (iia) using ruby coils and pod packing coils (pod pcs).During the procedure, the physician reported experiencing some resistance while re-positioning a ruby coil and that it unintentionally detached within the lantern delivery microcatheter (lantern).The lantern and the ruby coil were retracted from the patient and the ruby coil was successfully removed from the lantern.The physician continued the procedure and detached a pod pc into the target vessel using the same lantern.Subsequently, it was reported that the pod pc had migrated out of the target vessel into the patient's anatomy.The physician retrieved the pod pc using a snare, and the pod pc was removed from the patient.The procedure was completed using additional ruby coils.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key8634170
Report Number3005168196-2019-01001
Device Sequence Number16608780
Product Code HCG
UDI-Device Identifier00814548019086
UDI-Public00814548019086
Combination Product (Y/N)Y
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K173614
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/23/2019
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberRBY2C1240
Device Lot NumberF86591
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/23/2019
Initial Report FDA Received Date05/22/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age72 YR
Date Report Sent to FDA01/01/2005
-
-