| Model Number |
9004 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Reocclusion (1985)
|
| Date of Event |
02/25/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Analysis.The sample was not returned from the user facility; therefore, a device evaluation is unable to be performed.A lot history review revealed this is the only complaint associated with this lot.A review of the device history record (dhr) indicates the lot was manufactured to specification.Conclusion: the actual sample was not returned for evaluation.The dhr found nothing to indicate a manufacturing related cause for this event.The investigator assessed the event was possibly related to the study device or procedure.Based on the instructions for use (ifu), the occurrence of reocclusion is an inherent risk of any pta procedure, and has been reported in clinical trials of drug coated balloons.If additional information becomes known to the manufacturer, a supplemental report will be provided with all relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported through a post market clinical study that during the index procedure, a lutonix percutaneous transluminal drug coated balloon (dcb) dilatation catheter was used to treat the target lesion located in the left popliteal artery.Approximately 6 months after the index procedure, the patient¿s left popliteal artery was reportedly reoccluded via duplex ultrasound (dus) imaging.A revascularization involving a dcb was performed and health care professional (hcp) deemed it was successful.The investigator assessed that the event possibly related to the study device, but not related to the procedure.The sample was discarded by the user facility and is not available for evaluation.No adverse patient effects were reported.
|
| |
|
Search Alerts/Recalls
|