• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC 1.4MM X 12.8MM FRTR,CRNP; MOTOR, DRILL, ELECTRIC - CUTTING BURR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC 1.4MM X 12.8MM FRTR,CRNP; MOTOR, DRILL, ELECTRIC - CUTTING BURR Back to Search Results
Catalog Number P-CRN-G1
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/08/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Udi: (b)(4).Concomitant medical devices: 4 cutter devices.The actual device was returned for evaluation.Quality engineering evaluated the device and determined that the device had a broken tip.The external features of the returned device were visually inspected and observed that the tip of the cutter was broken.The cutter was examined and photographed using 28x magnification.Based on the photographs taken, the angle indicated that there was excessive force applied or an improper sized burr was being used.Therefore, the reported condition was confirmed.A review of the device history record was performed and no non-conformances were detected related to the reported condition.The assignable root cause was determined to be due to excessive lateral or side to side force or an improperly sized cutter was used which is improper device handling; user error.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Event or Problem Description
This is event 2 of 5 of the same event.It was reported that during a craniosynostosis surgical procedure, it was observed that five cutter devices snapped in half.According to the surgeon the cutter devices broke roughly fifteen minutes into use, without aggressively drilling.It was unknown if there were delays to the surgical procedure.There was patient involvement.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The actual device was returned for evaluation.Quality engineering evaluated the device and determined that the broken tip was confirmed.The external features of the returned cutter were visually inspected and it was observed that the tip of the cutter was broken.The cutter was examined and photographed using 28x magnification.Based on the photographs taken, the angle of the break indicated that there was excessive force applied or an improper sized cutter was being used.The root cause was due to excessive lateral or side to side force or improper sized cutter was being used.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to user error.A review of the device history record was performed and no non-conformance's were detected related to the reported condition.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.4MM X 12.8MM FRTR,CRNP
Common Device Name
MOTOR, DRILL, ELECTRIC - CUTTING BURR
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
MDR Report Key8634730
Report Number1045834-2019-53624
Device Sequence Number1941698
Product Code HBC
UDI-Device Identifier00845384012634
UDI-Public845384012634
Combination Product (Y/N)N
Initial Reporter StateDC
Initial Reporter CountryUS
PMA/510(K) Number
K011444
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberP-CRN-G1
Device Lot NumberM483134606
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/14/2019
Initial Date Received by Manufacturer 05/08/2019
Supplement Date Received by Manufacturer05/22/2019
Initial Report FDA Received Date05/22/2019
Supplement Report FDA Received Date05/31/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-