| Model Number |
G13287 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
05/09/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(6).This report includes information known at this time.A follow-up report will be submitted at the conclusion of the investigation or, when additional relevant information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported, a (b)(6) male was undergoing an inferior vena cave filter removal (cook celect platinum) when the filter got stuck (endothelialized in 3 months).The snare loop from the gunther tulip vena cava filter retrieval set "broke" after it was engaged with the filter hook.The filter was not able to be removed.There has been no effect on the patient; however part of the loop remained with the filter hook after it broke.
|
| |
|
Event or Problem Description
|
|
Additional information was received on (b)(6) 2019, noting that the physician has no plans to retrieve the segment portion of the loop.
|
| |
|
Additional Manufacturer Narrative
|
|
This report includes information known at this time.A follow-up report will be submitted at the conclusion of the investigation or, when additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Corrected fields: b5, d3, f14.The device associated with this report was manufactured by william cook europe, not cook incorporated.The complaint file will be transferred to william cook europe and cook, inc.Will not send any additional follow-up reports regarding this event.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
Correction: the device associated with this report was manufactured by william cook europe, not cook incorporated.The complaint file will be transferred to william cook europe and cook, inc.Will not send any additional follow-up reports regarding this event.
|
| |
|
Search Alerts/Recalls
|