Plant investigation: the actual device was returned to the manufacturer for physical evaluation.An exterior visual inspection of the returned cycler showed no signs of physical damage.A simulated treatment was performed and during dwell two the screen of the cycler permanently went blank.It was identified that the cause for the blank screen was due to an internal short on transformer (t1) on the inverter board.The inverter board is located on the rear of the touch screen.A known good inverter board was installed and the display became fully operational.The voltage check passed.A simulated treatment was performed and passed.A simulated treatment was performed and completed without failure.The cycler weighed fill volumes values were within tolerance.The cycler underwent and passed a system air leak test and valve actuation test.The load cell verification was within tolerance.An internal visual inspection of the returned cycler encountered no other discrepancies.A review of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, a device history record (dhr) review was performed and verified that the results of the in-progress and final quality control (qc) testing met all requirements.Upon completion of the evaluation, the reported issue was confirmed and the cause was determined to be an internal short on the transformer on the inverter board.The cycler was refurbished following the evaluation.
|
A peritoneal dialysis (pd) patient reported that they received a patient line is blocked message during drain.A review of the patient¿s treatment records identified that the patient drained 3291 ml during drain 5 of the treatment.This drain volume is 164% the patient's prescribed fill volume of 2000 ml.At that point in time, the technical support representative advised the patient to discontinue use of the cycler and to notify their peritoneal dialysis registered nurse (pdrn) of the event.A replacement cycler was issued to the patient.Additional information was requested, however to date has not been provided.Upon physical evaluation of the cycler by the manufacturer, a blank screen was encountered during dwell 2 of a simulated treatment.There was evidence of an internal short identified on the transformer on the inverter board.
|