• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEODENT - JJGC S.A. TI DRIVE ACQUA IMPLANT (3.3) 3.5X8 MM; ENDOSSEOUS DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEODENT - JJGC S.A. TI DRIVE ACQUA IMPLANT (3.3) 3.5X8 MM; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Catalog Number 140.752
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code Tissue Damage (2104)
Date of Event 03/21/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.Considering the surgical methodology, it was seen that the dentist has selected an implant design which is not recommended for the patient's bone quality.The investigation of the complaint was carried out considering the analysis of the information given by the dentist in the guarantee form, visual conference of the product returned by the dentist and the evaluation of relevant production records.
 
Event or Problem Description
(b)(4).The dentist reported that 5 months and 23 days after the dental implant was installed in intraoral region 29#, its non- osseointegration was observed.Moreover, 40n.Cm of primary stability was achieved and the patient presented bone type ii.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TI DRIVE ACQUA IMPLANT (3.3) 3.5X8 MM
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
NEODENT - JJGC S.A.
av. juscelino kubitschek de
oliveira, 3291 cic
curitiba PR 81270 -200
BR  81270-200
MDR Report Key8639760
Report Number3008261720-2019-02099
Device Sequence Number19290610
Product Code DZE
UDI-Device Identifier07899878017073
UDI-Public07899878017073
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K133592
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2019
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 05/24/2019
Report Date (Section F) 05/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/27/2020
Device Catalogue Number140.752
Device Lot Number140162
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/03/2019
Was the Report Sent to FDA? No
Distributor Facility Aware Date05/24/2019
Device Age12 MO
Event Location Other
Date Report to Manufacturer05/03/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/03/2019
Initial Report FDA Received Date05/24/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age51 YR
-
-