U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC MULTI-SIDEPORT DUAL CHECK VALVE CATHETER INFUSION SET; DQO CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC MULTI-SIDEPORT DUAL CHECK VALVE CATHETER INFUSION SET; DQO CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number G06425
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 04/04/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Occupation: non-healthcare professional.Pma/510(k) number = pre-amendment.This report includes information known at this time.A follow-up report will be submitted at the conclusion of the investigation or, when additional relevant information becomes available.
 
Event or Problem Description
It was reported, upon receipt of a multi-sideport dual check valve catheter infusion set at the distribution center, a piece of hair was observed inside of the sealed, primary packaging.This observation was made upon incoming inspection of the device.It was not used for patient care.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Investigation - evaluation a review of the complaints history, device history record, quality control, and instructions for use (ifu) of the device was conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record shows no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.The product ifu instructs the user the user no to use the product if there if doubt as to whether the product is sterile an advises the user to inspect the product to ensure no damage has occurred upon removal from package.Based on the information provided, no product returned and the results of our investigation, it has been concluded the cause of this event is manufacturing deficiency related.Appropriate measures have been initiated to address this failure mode.We will continue to monitor for similar complaints.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MULTI-SIDEPORT DUAL CHECK VALVE CATHETER INFUSION SET
Common Device Name
DQO CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key8640245
Report Number1820334-2019-01286
Device Sequence Number2017016
Product Code DQO
UDI-Device Identifier00827002064256
UDI-Public(01)00827002064256(17)211207(10)9205706
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/07/2021
Device Model NumberG06425
Device Catalogue NumberMCIS-5.0-35-130-15.0
Device Lot Number9205706
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/21/2019
Supplement Date Received by Manufacturer05/30/2019
Initial Report FDA Received Date05/24/2019
Supplement Report FDA Received Date05/31/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-