It was reported, after inserting the picc line into the patient, they were not able to flush through the catheter.They discarded the product immediately and used a another picc line to complete the procedure.According to the information provided, there were no adverse effects to the patient due to this occurrence.
|
|
Investigation ¿ evaluation: a review of the complaint history, device history record, instructions for use (ifu), quality control, and specifications of the device, were conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of relevant manufacturing documents was conducted.It was concluded that the device aspect in question was visually/functionally inspected by quality control and no related gaps in production or processing controls were noted.The risk specification for this product includes the loss of use failure mode and identifies the risk controls that are in place to mitigate the risk of this type of failure.A software search revealed this complaint to be the only one associated with the complaint lot number.Based on the information provided, no product returned and the results of the investigation, a definitive root cause could not be determined.Per the quality engineering risk assessment no further action is required.Cook will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|