PENUMBRA, INC. LANTERN DELIVERY MICROCATHETER; DQY Back to Search Results
Catalog Number PXSLIMLAN150T90
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/25/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The product lot number was not provided, therefore, the manufacturing records could not be reviewed.This report is associated with mfr report numbers: 3005168196-2019-01031, 3005168196-2019-01032, 3005168196-2019-01033.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the renal artery using ruby coils and a lantern delivery microcatheter (lantern).During the procedure, the physician successfully placed a ruby coil using a lantern.The physician then felt resistance while advancing another ruby coil and, therefore, the ruby coil was removed.Upon removal, the physician reported that the coil was "stretched".The physician then switched to a new lantern and attempted to place a new ruby coil, however, the physician was unable to advance this ruby coil as well.Therefore, the ruby coil was removed, and the procedure ended at this point, despite the low packing density.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LANTERN DELIVERY MICROCATHETER
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key8640721
Report Number3005168196-2019-01036
Device Sequence Number11266873
Product Code DQY
UDI-Device Identifier00814548016696
UDI-Public00814548016696
Combination Product (Y/N)Y
Initial Reporter CountryGM
PMA/510(K) Number
K152840
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/25/2019
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberPXSLIMLAN150T90
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2019
Initial Report FDA Received Date05/24/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age71 YR
Date Report Sent to FDA01/01/2005
-
-