Catalog Number 71335554
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 01/01/1901
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that recurrent dislocations on (b)(6) 2017 & (b)(6) 2018 occurred.
 
Additional Manufacturer Narrative
The associated complaint devices were not returned.The clinical/medical team concluded, the doctor¿s statement that the patient¿s need for a 3rd revision, to ¿increase the head size and/or mobility¿and without a revision attempt, that certainly she will dislocate at some point again¿ indicates that the likely root cause for her multiple post-revision dislocations is due to an inappropriate head size, or another reason for unexpected joint laxity.To date, there is no indication that the patient has had the 3rd revision performed.No further clinical assessment is warranted at this time.A review of the production documentation for the corresponding products did not reveal any deviation from the standard manufacturing processes.A review of complaint history for the listed parts revealed no prior complaints for the listed batches.Without the actual device involved our investigation cannot proceed.If the device or new information is received in the future, the complaints can be re-opened.No further actions are being taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 3 HOLE ACET SHELL 54MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8642581
Report Number1020279-2019-02037
Device Sequence Number11173647
Product Code MBL
UDI-Device Identifier03596010598257
UDI-Public03596010598257
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K070756
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number71335554
Device Lot Number17BM17380
Initial Date Received by Manufacturer 05/03/2019
Supplement Date Received by Manufacturer05/03/2019
Initial Report FDA Received Date05/24/2019
Supplement Report FDA Received Date11/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age65 YR
-
-