| Model Number |
MMT-523RNAL |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/21/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that the insulin pump had motor error and motor position encoder error.The customer blood glucose level was 347 mg/dl and 262 mg/dl.Customer reports the drive support cap appears normal.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed rewind test, basic occlusion test, prime test and displacement test.However, device received stuck in the motor error loop during bolus/basal delivery and e70 alarm confirmed in the history file.The motor may have had an intermittent failure that was not detected during our testing.The motor was tested outside the device and passed.
|
| |
|