| Model Number |
DIPLOMAT |
| Medical Device Problem Codes |
Mechanical Problem (1384); Unintended System Motion (1430); Use of Device Problem (1670); Improper or Incorrect Procedure or Method (2017); Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923); Unintended Movement (3026)
|
| Health Effect - Clinical Codes |
No Code Available (3191); Missing Value Reason (3192)
|
| Date of Event |
05/09/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Summary: the complaint of the tulip and set screw is not justified.The claimed tulip and screw have no functional or structural deficits.A secure attachment of the rod is possible with the product without restriction.There are no additional, previously unknown risks.Due to the existing findings loosening of the set screw is a result of mishandling during the first implantation surgery.
|
| |
|
Event or Problem Description
|
|
Post op images of the patient showed that a set screw had loosened from the tulip and could be seen next to the screw / rod combination.Due to the loosened locking screw, a revision surgery of 2 hours duration was required after one week.For reasons of data protection, the clinic did not provide radiographs and patient data.
|
| |
|