• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK IRELAND LTD ECHOTIP PROCORE ENDOBRONCHIAL HD BIOPSY NEEDLE; FCG KIT, NEEDLE, BIOPSY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK IRELAND LTD ECHOTIP PROCORE ENDOBRONCHIAL HD BIOPSY NEEDLE; FCG KIT, NEEDLE, BIOPSY Back to Search Results
Model Number G34281
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Foreign Body In Patient (2687); No Known Impact Or Consequence To Patient (2692)
Date of Event 04/29/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Pma/510(k) #k160229.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4) cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195.Importer site establishment registration number: 3005580113.Investigation is still pending.A follow up mdr will be submitted to include the investigation conclusions.
 
Event or Problem Description
As reported to customer relations: "tip of the needle broke." dm unsure at time of initial reporting how much broke off, how many pieces, if any remains in patient, at what point of procedure the break occurred, if the patient has been harmed, or any other details.
 
Additional Manufacturer Narrative
Pma/510(k) #k160229.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: ed sutkowski cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195.Importer site establishment registration number: 3005580113.1 unit of lot c1547037 of echo-hd-22-ebus-o-c was returned opened in its original packaging the device related to this occurrence underwent a laboratory evaluation.The needle was found to be broken.Prior to distribution, all echo-hd-22-ebus-o-c devices are subjected to functional checks and visual inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.A review of the manufacturing records for echo-hd-22-ebus-o-c of lot number c1547037 did not reveal any discrepancies that could have contributed to this complaint issue.The review of relevant manufacturing records, confirms the failure mode has not previously occurred with the current lot number.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number c1547037.The notes section of the instructions for use, which accompanies this device instructs the user to; "visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use" and ¿this device is intended for use with an olympus ebus scope." clarification was requested as to the type of scope used and it was confirmed that a fuji eb-530us scope was used.As this did not meet the ifu requirement a new pr was requested to be opened.There is no evidence to suggest that the customer did not follow the instructions for use.The failure of needle broken was observed in the laboratory.A definitive root cause for the customer complaint could not be determined as circumstances of use cannot be replicated in the laboratory.A possible root cause could be attributed to the device being used in a flexed or twisted position as indicated in the additional information.Complaint is confirmed as the failure was verified in the laboratory.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Event or Problem Description
As reported to customer relations: "tip of the needle broke." dm unsure at time of initial reporting how much broke off, how many pieces, if any remains in patient, at what point of procedure the break occurred, if the patient has been harmed or any other details.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHOTIP PROCORE ENDOBRONCHIAL HD BIOPSY NEEDLE
Common Device Name
FCG KIT, NEEDLE, BIOPSY
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
MDR Report Key8644181
Report Number3001845648-2019-00223
Device Sequence Number2129835
Product Code FCG
UDI-Device Identifier00827002342811
UDI-Public(01)00827002342811(17)211008(10)C1547037
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/08/2021
Device Model NumberG34281
Device Catalogue NumberECHO-HD-22-EBUS-O-C
Device Lot NumberC1547037
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/28/2019
Was the Report Sent to FDA? No
Distributor Facility Aware Date05/29/2019
Event Location Hospital
Initial Date Received by Manufacturer 04/29/2019
Supplement Date Received by Manufacturer04/29/2019
Initial Report FDA Received Date05/27/2019
Supplement Report FDA Received Date06/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-