| Model Number |
33103 |
| Medical Device Problem Code |
Nonstandard Device (1420)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the male external catheter had too strong of an adhesive.No patient injury reported.
|
| |
|
Event or Problem Description
|
|
It was reported that the male external catheter had too strong of an adhesive.No patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was unconfirmed.Visual evaluation noted 15 unopened ultraflex mecs received.No obvious defects were noted.An adhesive peel test was performed and the samples were cut to the required width (0.70" +/- 0.05").The adhesive peel strength was found to be within specification (0.60-2.10lbf).The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿description/indication the self-adhering male external catheter is designed for the management of male urinary incontinence.Contraindication do not use on irritated or compromised skin.Precaution do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Directions: to apply 1) wash penis with mild soap and warm water.Dry thoroughly.2) trim pubic hair if necessary.3) unroll self-adhering catheter over penis.4) gently squeeze the catheter to properly seal adhesive to the skin.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.5) connect to drainage device.Directions: to remove gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive.".
|
| |
|
Search Alerts/Recalls
|