STRYKER ENDOSCOPY-SAN JOSE PKG., TOMCAT SHAVER BLADE, F - SERIES - 5.5MM; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL Back to Search Results
Catalog Number 0375565000
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/09/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that the device broke during the procedure.All pieces were retrieved and the procedure was completed successfully.
 
Additional Manufacturer Narrative
Alleged failure: during case, shaver cutter tip broke off, about 10mms worth.Surgeon able to retrieve.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be blade comes in contact with a hard object causing damage to the teeth and hitting the housing edges.Excessive pressure, such as bending or prying, may cause the arthroscopic shaver blades to bend/break.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
 
Event or Problem Description
It was reported that the device broke during the procedure.All pieces were retrieved and the procedure was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG., TOMCAT SHAVER BLADE, F - SERIES - 5.5MM
Common Device Name
BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key8647087
Report Number0002936485-2019-00219
Device Sequence Number263385
Product Code GFA
UDI-Device Identifier37613327061933
UDI-Public37613327061933
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0375565000
Device Lot Number18156CE2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/20/2019
Initial Date Received by Manufacturer 05/10/2019
Supplement Date Received by Manufacturer05/10/2019
Initial Report FDA Received Date05/28/2019
Supplement Report FDA Received Date08/12/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-