• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THORATEC SWITZERLAND GMBH CENTRIMAG MOTOR, US

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC SWITZERLAND GMBH CENTRIMAG MOTOR, US Back to Search Results
Model Number 102956
Device Problems Pumping Stopped (1503); Vibration (1674); Noise, Audible (3273)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/09/2019
Event Type  Injury  
Manufacturer Narrative
Patient identifier, sex: the patient initial and gender was not provided. Description of problem or event: this is reported again the cmag motor. The console is reported under mfr report # 2916596-2019-02312. No further information was provided. A supplemental report will be submitted when the manufacturer¿s investigation is completed.
 
Event Description
It was reported that patient was on centrimag (cmag) support during a surgical procedure that was unrelated to the patient being placed on cmag. Towards the end of the procedure, a humming and clicking noise was heard coming from the cmag motor. The clinical team observed the rpms drop from 4300 to 3250 and the flow waveform on the cmag monitor dropped to 0 sounding an alarm. The system was stopped. The cmag system was replaced with a backup and the system was initiated again.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCENTRIMAG MOTOR, US
Type of DeviceCENTRIMAG MOTOR
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ CH-8005
Manufacturer (Section G)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ CH-8005
Manufacturer Contact
bob fryc
6035 stoneridge drive
pleasanton, CA 94588
7818528204
MDR Report Key8647205
MDR Text Key146274509
Report Number2916596-2019-02311
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 08/28/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received05/28/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number102956
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received08/08/2019
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/23/2018
Is the Device Single Use? No Answer Provided
Is This a Reprocessed and Reused Single-Use Device?
Type of Device Usage
Removal/Correction NumberFA-Q319-MCS-1

-
-