Product complaint # (b)(4).The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Evaluation: five unopened samples of product code j269, lot mgz159 were returned for analysis.During the visual inspection of five samples, no defects were found on the packages.The samples were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strands and no defects were observed.A functional test was performed using an instron and the pull force were above the minimum requirements.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the condition of the samples, no performance pull off suture needle was found, and the tested samples met the finished goods requirements and the reported complaint could not be observed.If further details are received at a later date a supplemental medwatch will be sent.
|