Upon receipt at medtronic¿s quality laboratory, visual examination revealed the holder was received attached to the band.No damages were noted on the holder.Both retention sutures were intact.The band met the size dimensions for a size 30mm band.Radiograph revealed no evidence of distortion and / or fractures along the band.Conclusion: the ring met the size dimensions for a size 30mm cg future ring.Radiography revealed no evidence of distortion and / or fractures along the band.The band was determined to be acceptable.There are times when valve repair is attempted using a repair device and subsequent post-repair evaluation demonstrates an inadequate result.In this case ,the ring was replaced due to a sizing issue with a different size and model ring.Updated coding in h6.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|