|
The concerto coil was returned for analysis.The concert coil was found to be detached from the pushwire.The concerto implant coil and pushwire were then decontaminated.No bends or kinks were found with the coil¿s pushwire.The coin was found against the lumen stop.The coil shied was found intact.The concerto implant coil was found damaged and had lost its original shape.No other anomalies were observed.Based on the device analysis and reported information, the report of ¿poor lay up or framing¿ could not be confirmed.Possible causes of failure include patient vessel tortuosity, catheter kick back, high blood flow, or incorrect sizing of implant.The patient¿s vasculature was minimal in tortuosity, the blood flow was normal, and the catheter was not under stress.It¿s possible the damage to the implant coil contributed to the event; however, the cause for the event could not be determined.Per the concerto coil instructions for use (ifu): warning: "if the coil does not move with a one-to-one motion, or repositioning is difficult, the coil has been stretched and could possibly break.Gently remove and discard both the catheter and coil." information is provided in the future, a supplemental report will be issued.
|