The lay user/patient's test strips have been returned and evaluated by lifescan product analysis with the following findings: the test strips failed testing; the reported issue was confirmed.The test strips were found to have been sticking together.This could be due to static or small amounts of adhesive.During visual inspection of the returned test strips, the strips come apart easily when handled.Complaints relating to the reported product were evaluated.It was concluded that the number of complaints for the product did not breach thresholds indicative of a systemic issue.On (b)(6) 2019, the lay user/patient contacted lifescan (b)(4), alleging that their ot verio test strips were sticking together.This complaint was initially ruled out because during customer services troubleshooting, there was no indication to reasonably suggest that the product malfunctioned and there was also no allegation of an adverse event as a result of the reported issue.
|