• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTERVASCULAR SAS INTERGARD HEMABRIDGE WOVEN; VASCULAR POLYESTER GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTERVASCULAR SAS INTERGARD HEMABRIDGE WOVEN; VASCULAR POLYESTER GRAFT Back to Search Results
Model Number HEW3010BRIDGE
Medical Device Problem Code Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Codes Death (1802); Blood Loss (2597)
Date of Event 04/21/2019
Type of Reportable Event Death
Additional Manufacturer Narrative
A review of the complaint device history records is ongoing, results are pending.The investigation is still ongoing.A follow-up report will be sent upon completion of the investigation.
 
Event or Problem Description
During a total arch replacement and tevar procedure, the surgeon connected the involved graft with a 3-branch hemashield graft, and the operation was completed.After angiostomy, bleeding was identified on the proximal part of the graft where a needle puncture was performed for deairing.At the completion of this first operation, small bleeding was observed but the surgeon thought it would stop and closed the patient's chest.However, the patient's condition was not good, and a second operation was performed.On the second exploration, bleeding occurred in the area where the two grafts were connected.Various hemostatic agents, like flosea and surgical berilast, were applied, which reduced the bleeding amount.About x20 pledgeted reinforcement interrupted sutures were placed on the blood oozing surface of the graft, and finally bleeding stopped.Blood transfusion and hemostatic pledget repair were performed for 2.5 hours.However, a few hours later, the patient had a heart arrest.Ecmo was performed but the patient died in a week (exact date of death is unknown).Additional information: transfusion : rbc 320cc 30ea, ffp 40cc 16ea.The patient had no bleeding tendency before the operation.
 
Additional Manufacturer Narrative
Expiration has been corrected.(4117) device is not accessible for testing as it remained implanted in the patient.(3331/213) a review of the complaint device history records, indicated that the graft was processed and inspected according to established procedures and was therefore released following acceptable quality inspections and tests.Moreover, the review of the water permeability testing records of products coated on the same day as the complaint device indicated values well within product specifications (< 5 ml/cm²/min).(4109/213) the review of post-marketing historical data indicated that no other similar complaint was reported for the same lot number.(4315) no conclusion can be drawn since the product remained implanted.However, the conducted investigation suggests that the product was not defective at the time of manufacturing.The case has been reviewed by our corporate medical officer, his assessment is as follows: "it is difficult to speculate on the possible causes of the bleeding.It is however very uncommon for volume and area of interest.It could be assumed that due to the needle insertion and the subsequent closure the area surrounding the hole was subjected to additional stretch and trauma.Therefore, the collagen coating could have been disrupted causing the diffuse bleeding.However this remains speculation as in absence the specimen for analysis it is impossible to draw a conclusive assessment of the causes of the event.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERGARD HEMABRIDGE WOVEN
Common Device Name
VASCULAR POLYESTER GRAFT
Manufacturer (Section D)
INTERVASCULAR SAS
z.i. athelia i
la ciotat cedex, 13705
FR  13705
MDR Report Key8649086
Report Number1640201-2019-00030
Device Sequence Number243192
Product Code DSY
UDI-Device Identifier00384401001027
UDI-Public00384401001027
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K013651
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date05/31/2022
Device Model NumberHEW3010BRIDGE
Device Catalogue NumberHEW3010BRIDGE
Device Lot Number17F01
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/01/2019
Supplement Date Received by Manufacturer06/07/2019
Initial Report FDA Received Date05/29/2019
Supplement Report FDA Received Date07/05/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death; Required Intervention;
Patient Age68 YR
Patient Weight63
-
-