| Model Number |
694775 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
04/27/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient had an infection.The organism was identified as enterococcus faecalis bacteremia.A transesophageal echocardiogram (tee) noted artifact verse vegetation on the right atrial lead.Antibiotic treatment was necessary.The device and leads were explanted and replaced with a leadless implantable pulse generator (ipg).No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary : the full lead was returned and analyzed, no anomalies were found.Visual analysis of the lead indicated apparent explant damage.If information is provided in the future, a supplemental report will be issued.
|
| |
|